A Pitot tube lying on the wall reads the wall's friction
Worth reading first: The instrument in the answer · The layer with no length in it.
Every instrument met so far in this subject has inferred something about a flow from something easier to read. The price of knowing the flow rate is an orifice plate that turns a flow rate into a pressure difference and charges a permanent loss for doing so. The instrument in the answer is the way a calibration constant hides the model a meter was built on, and a magnetic meter turns out to be blind to whole families of distorted profiles. All of them measure how much fluid is moving. None of them measures the force the fluid exerts along the wall, and that is the quantity the drag of a ship, the pressure drop of a pipeline and the heat transfer of an exchanger are all built on.
The difficulty is not subtle. The wall shear stress acts on a skin, and the skin cannot be weighed separately from the plate it covers. A floating element — a small patch of wall cut free and mounted on a balance — does it directly, and it is a laboratory instrument of great delicacy: a gap around the patch disturbs the flow it is trying to measure, the force is tiny and the pressure gradient pushes on the patch’s edges. In 1954 J. H. Preston, at Cambridge, suggested doing it with a Pitot tube, a round hypodermic tube resting on the surface and facing the flow. It reads a pressure, which is easy. The argument that the pressure is the friction is a piece of dimensional reasoning, and following it through says both why it works and where it fails.
The inner layer has nothing else to depend on
The argument rests on the law with no length in it. Close to the wall of a turbulent flow, the mean speed at a height is set by the wall shear stress , the density , the viscosity and the height, and by nothing about the flow further out — not the free stream’s speed, not the pipe’s diameter, not the pressure gradient, not the history of the layer. There is one velocity that can be made out of the first three, , and one length, , so the law must take the form
written . Very near the wall it is ; further out it becomes the logarithm, and between them is a buffer layer joining the two. Spalding wrote a single formula in 1961 that runs smoothly through all three, and that is the formula used here, with the logarithm’s constants and .
Now put a round tube of outer diameter on the wall, facing upstream. It reads a dynamic pressure, , the excess of the pressure in its mouth over the static pressure at the wall. If the tube is small enough to sit inside the inner layer, then the speeds across its mouth are fixed by the inner law, and the pressure it reads is fixed by the speeds, the density and the tube’s diameter. So depends on , , and , and only those. Four quantities in three dimensions make one dimensionless relation:
That is the whole of Preston’s idea. The function is the same for every tube of the same shape on every surface where the inner law holds. It can be found once, in a pipe, where the wall stress is known exactly from the pressure drop, and then carried to an aircraft wing or a ship’s hull. The logarithms of the two groups are conventionally written and , and the standard calibration is V. C. Patel’s of 1965, three formulas for three ranges of tube size.
Reproducing the calibration from the inner law
If the tube read the speed at its own centre, the calibration would follow from the inner law with no further constant: a dynamic pressure at . Doing that computation shows it is wrong, and wrong by a lot. In the viscous sublayer, where exactly, reading at the centre gives , the second number being half the logarithm of two. Patel’s calibration for small tubes is . The difference in the intercept is 0.11 in a logarithm of the stress, a factor of 1.30 in the inferred friction. A tube interpreted as reading at its centre infers a wall stress 23 per cent too low.
The reason is that the speed varies across the tube’s mouth, and a Pitot tube in a shear flow does not read the speed at its geometric centre. The fast fluid at the top of the mouth is brought to rest with more total pressure than the slow fluid at the bottom, the stagnation streamline bends downward towards the slow side, and the pressure in the mouth is that of a streamline further from the wall than the centre. B. G. MacMillan measured the shift in 1956 and found it close to a constant fraction of the diameter: the effective centre sits about 0.15 above the geometric one.
Putting MacMillan’s shift into the calculation — reading the inner law at instead of — moves the computed intercept from 0.1505 to 0.0367 and lands on Patel’s 0.037 almost exactly. That is a check worth having, because the shift was measured in one experiment and the calibration in another, eight years apart, and the inner law connects them with nothing adjustable. In the logarithmic range, at the other end of the calibration, the same displaced reading is within 1.1 per cent of Patel’s formula for every tube size. The two ends of a calibration fitted to measurements are reproduced by a law of the wall and a displacement, and neither was fitted to the other.
The middle is not reproduced. In the buffer layer, for tubes between about 15 and 60 wall units across, the model infers a friction up to 11.8 per cent lower than Patel’s calibration does, the worst at , a tube about 35 wall units in diameter. That is not noise. Either Spalding’s buffer-layer profile is too slow there, or the effective centre is not a fixed fraction of the diameter when the tube spans the region where the profile bends — or both, and the measurements do not separate them. It is the region where the velocity profile’s curvature is largest, and a displacement correction derived for a linear shear has no business being exact in it. That is exactly why Patel’s calibration is empirical in the middle: the dimensional argument guarantees that one curve exists, and only a measurement says what shape it has where the inner law is least simple.
What the tube sees
The figure below shows the inner law itself, on a logarithmic axis of height, with the tube drawn on it. The mouth spans everything from the wall to its own diameter. What it reports is one point, the speed at its effective centre.
Two consequences follow from the picture. The first is that a fat tube and a thin tube on the same wall read different pressures and infer the same friction, because they read different points of one known curve. That is the experimental signature of the inner law holding, and Preston used it: four tubes of different sizes on one pipe wall gave one stress. Where the inner law fails, the inferences disagree, and the disagreement is the warning.
The second is that a tube cannot be made arbitrarily large. Once its effective centre reaches the part of the layer that does feel the outer flow, the reading is no longer a function of the friction alone, and the calibration breaks. For a pipe that boundary is near the pipe’s centreline; for a boundary layer it is the wake, the outer part where the profile rises above the logarithm by an amount that depends on the layer’s history and its pressure gradient.
How hard the friction answers the pressure
A measurement’s value lies in how an error in what is read becomes an error in what is inferred. Along the calibration curve the friction goes as a power of the reading, , and the exponent changes with the tube’s size.
In the sublayer it is exactly a half. The speed there is proportional to the friction times the height, so the dynamic pressure goes as the square of the friction, and the friction as the square root of the pressure. A small tube therefore halves every error in the pressure reading: a one per cent error in becomes half a per cent in . That is a generous exchange, and it is paid for in a pressure that is tiny — a hypodermic tube a fraction of a millimetre across on a water tunnel’s wall reads a few pascals.
Further out the exponent rises. At a diameter of 300 wall units it is 0.88, and at 2,000 it is 0.90 and still climbing. In the logarithmic layer the speed goes as the square root of the friction times a logarithm that itself contains the friction, so the pressure is nearly linear in and almost none of the error is absorbed. A large tube buys a comfortable pressure to read at the price of passing its error almost straight through.
Where the outer flow gets in
The inner law knows nothing about the outer flow by construction. A real boundary layer has an outer flow, and the departure of the mean profile from the logarithm in its outer part is described by Coles’s wake function, a smooth rise of total height across the layer. is about 0.55 on a flat plate with no pressure gradient and grows without limit as an adverse gradient pushes the layer towards separation. A tube whose effective centre reaches the wake reads the wake’s extra speed as extra friction.
On a flat plate the error is small even for a large tube: a tube a tenth of the layer thick reads 0.3 per cent high, and a fifth 1.0 per cent. Near separation, with , the same tubes read 2.0 and 7.4 per cent high. The rule that follows is the one practitioners use: the tube’s diameter should be a small fraction of the layer, and the smaller the fraction the stronger the adverse gradient. Patel’s own tests put the limit in a pressure gradient as a condition on a dimensionless gradient, , and Preston tubes are known to fail in exactly the flows where the friction matters most — the layer that is about to separate, whose friction is going to zero while its outer flow is doing something else.
There is a related way the inner law fails that the figure does not show. On a rough wall the inner law changes: the logarithm is shifted down by an amount that depends on the roughness height in wall units, as a second length at the wall describes, and a calibration made on a smooth pipe misreads it. And in a layer whose friction is changing rapidly along the surface the inner law takes a few layer thicknesses to catch up, so a tube just downstream of a step change reads the friction of the surface upstream.
The checks
The model’s agreement with Patel’s calibration was computed rather than eyeballed. Spalding’s formula was inverted by bisection, and inverting the result returned the starting height to . The sublayer exponent was obtained by a finite difference along the model’s own curve and is 0.500 to three figures. The undisplaced intercept is 0.1505, which is as the algebra says, and the displaced one is 0.0367.
What the picture cannot show
Every number here comes from a one-dimensional caricature of a three-dimensional instrument. The tube is treated as reading the dynamic pressure of one streamline, displaced by a constant fraction of its diameter. In reality the tube is a blunt body sitting on a wall in a shear flow; it produces its own horseshoe vortex wrapped around its base, and the pressure in its mouth is set by a local flow that no inner law describes. The constant displacement summarises that flow for small tubes and for large ones, and the buffer-layer gap is where the summary stops being good. Nothing in the figures predicts how the displacement varies with the tube’s wall thickness or the shape of its mouth, and Patel found both matter.
Nor do the figures say anything about compressibility. On a supersonic wind tunnel’s wall the density varies across the layer, the inner law has to be transformed before it applies, and the tube itself may sit behind a shock. And the static pressure the tube’s reading is referred to is measured at a wall tapping next to it, an instrument with its own error of a size comparable to the dynamic pressure of a small tube.
The convention the calibration is written in
The two groups are not the obvious ones. Preston’s argument produces and ; Patel divided both by four, so that the radius rather than the diameter appears, and took their common logarithms. A calibration quoted as against and used with the factor of four forgotten is wrong by 0.60 in each logarithm, and since the calibration is not a straight line the error does not cancel. The stress is the mean wall shear stress, and the pressure is the tube’s reading minus the wall’s static pressure at the same streamwise station, not the free stream’s.
And the inner law is assumed rather than derived. Its form is a dimensional argument, and its constants, here and , are measured. Where the constants come from, and why they are the same in a pipe and on a plate, is a separate question; the Preston tube is one of the instruments that showed they were.
Who measured it
Preston published “The determination of turbulent skin friction by means of Pitot tubes” in the Journal of the Royal Aeronautical Society in 1954, after calibrating four tubes in a pipe and using them on a flat plate. MacMillan’s displacement correction appeared in 1956 in an Aeronautical Research Council report. Patel’s 1965 paper in the Journal of Fluid Mechanics recalibrated the tube over a wider range with more care, gave the three formulas used here and the conditions on the pressure gradient, and remains the standard. The instrument is still used, because it is cheap, robust and easy to move along a surface, and because no other instrument gives the friction as a pressure.
The idea underneath it is older and simpler than the instrument. It is the observation that a profile a meter cannot see in the core of a pipe is fixed near the wall by one number, and that a gauge sitting where the profile is universal reads that number whatever is happening elsewhere. The airspeed indicator on an aeroplane makes the mirror-image bet: its Pitot sits far from every wall, where the flow knows nothing of the friction, and reads the speed.
Still open: whether a tube can read a layer that is about to separate
The calibration fails in a strong adverse gradient because the inner law fails there, and the inner law fails because the friction is going to zero while the pressure gradient is not. Near separation the length grows without bound and the scales that matter become and the pressure gradient itself: a velocity and a length . Near-wall similarity in those variables is due to Stratford, who used it to design a pressure recovery that holds a layer on the edge of separation along its whole length.
The calculation that would follow writes the tube’s reading in those variables instead of in wall units and asks whether a second universal curve exists — a calibration in which the reading is a function of the pressure gradient and a stress that is now a small correction. If it does, a pair of tubes of different diameters at one station gives two equations for two unknowns, the stress and the gradient, and the Preston tube becomes an instrument for the one layer it currently cannot measure. Whether the near-separation profile is universal enough to support that is a question the measurements of attached-but-failing layers could answer without a new experiment.
Shares its objects with
Essays naming at least two of the same things, that neither author linked.
- A grain moves when its drag tips it out of its pocket — both name log law, measurement, model limit, wall shear
- The Obukhov length is where buoyancy takes over from the wind — both name dimensional analysis, log law, measurement, model limit
- The paint that measures the wrong field — both name calibration, instrument, measurement, model limit
- Weighing what is missing — both name instrument, measurement, model limit, total pressure